姚晓华,王丹,刘博姬,等. 剪切波速度成像定量鉴别乳腺良恶性病灶的临床研究[J]. 肿瘤影像学, 2017, 26(4): 272-278
目的:
探讨一种最新剪切波弹性成像技术剪切波速度成像(shear wave speed imaging,SWSI)对乳腺良恶性病灶的鉴别价值。
方法:
选取109例于上海市第十人民医院经手术病理证实的女性患者乳腺病灶共109个。术前乳腺病灶均行普通超声及剪切波弹性成像检查,测得杨氏弹性模量和速度模量,包括平均值(Emean)、SD及以正常乳腺腺体为参照的硬度比值(ratio)。根据病灶最大径,将病灶分为10 mm、>10~20 mm、>20 mm三组,以手术病理结果为金标准,绘制受试者工作特征(receiver operating characteristic,ROC)曲线,获取弹性参数,分析其鉴别良恶性病灶的价值。
结果:
109个病灶中,良性85个,恶性24个。比较不同组内鉴别良恶性病灶的弹性成像参数,除10 mm组外,其余组中恶性病灶的杨氏弹性模量及剪切波速度均高于良性病灶,差异有统计学意义(
P
<0.001)。比较所有参数在各组内鉴别良恶性病灶的诊断效能,Emean优于SD及ratio,差异均有统计学意义(
P
<0.01)。其中>20 mm组曲线下面积(area under curve,AUC)及诊断效能较其他两组高。以弹性模量及速度的平均值39.20 kPa、3.51 m/s作为截断值,鉴别乳腺病灶良恶性的灵敏度、特异度分别为86.7%、100%,86.7%、100%,差异均有统计学意义(
P
0.001),但组间杨氏弹性模量及速度差异无统计学意义(
P
>0.05)。
结论:
SWSI有助于鉴别乳腺病灶良恶性,其弹性参数Emean、SD及ratio为乳腺疾病提供了更多诊断依据。
Objective:
The aim of this study was to assess the value of a new technique of shear wave speed (SWS) imaging in quantitativy differentiating benign from malignant breast lesions.
Methods:
109 breast lesions in109 patients who underwent routine ultrasonography (US) and SWS imaging before surgical excision were analyzed. The quantitative
SWS imaging parameters including: Mean and SD of shear wave speed (m/s) and elasticity modulus ( kPa) value and ratio referenced as normal breast tissue were obtained
respectively. The lesions were categorized into three groups according to size (group 1
10 mm; group 2
10-20 mm; and group 3
20 mm). According to pathologic results
ROC curve was drawn to evaluate the best cut-off value in differentiating benign from malignant lesions. Diagnostic performances of SWS imaging parameters were calculated and compared.
Results:
There were 85 benign nodules and 24malignant nodules of the 109 breast lesions. The elastic values of malignant nodules were significant higher than those of benign one(all
P
0.001)
except for group1(10 mm). For diagnostic performance in all groups
the values of E-mean were higher than SD and ratio
(all
P
0.01).On ROC curve
the area under the curve (AUC) of group 3 were higher than group 1 and 2. No significant differences of AUC between SWS imaging parameters were found (all
P
0.05). Taking 39.20 kPa and 3.51 m/s as the cut-off values
sensitivity of 86.7% and 86.7%
specificity of 100% and 100% were achieved ( all
P
0.001). No significant differences of SWS imaging parameters between Young's elastic modulus and shear wave speed (SWS) were found (all
P
0.05).
Conclusion:
The new technique of shear imaging is helpful for the differential diagnosis of breast lesions.